AseptiSac™ AMC
MDI AseptiSac™ AMC autoclavable beta bags with chute are designed for transfer of sterile components into isolators or removal of aseptic waste out of isolators. These bags can withstand 2 autoclave cycles at 123°C for 30 minutes each.
The beta port of AseptiSac™ AMC autoclavable beta bags fits onto alpha port on isolators, while maintaining the sterile contamination free environment.
- Sterile components into isolators or removal of aseptic waste
- Removable/reusable handle system
- Fast and easy contamination free transfer
- Increased flexibility and reduced manufacturing time
- Minimize validation costs
- No cleaning validation
- Minimized manual intervention
| Construction | |
| Flange and Cover | High temperature polycarbonate |
| Beta Port | Platinum cured silicone |
| O-Ring | Platinum cured silicone |
| Weld Ring | High density polyethylene (HDPE) |
| Bag layers + Chute | Tyvek |
| Load Bearing Capacity | 25 kg |
| Specifications | |
| Sizes Available | Beta Port: 190 mm Beta Bag: 25 liter* *Please contact for customized sizes |
| Assurance | |
| Bacterial Endotoxin | Aqueous extracts exhibit <0.25 EU/ml as established by Limulus Amoebocyte Lysate (LAL) test as per USP <85> |
| Fiber Release | Passes test as per USP and comply with USFDA Title 21 CFR Part 210.3(b)(6) for fiber release |
| Particle Release | The filtrate complies with USP <788> test for particulate matter in injections |
| Bioburden Levels | Bioburden level is < 1000 cfu/bag as per ISO 11737-1 |
| File Type | Download |
|---|---|
| Datasheet | Download |