UA-44035381-1
  • AcuFlo™ Accurate Pump PCS Tubing

AcuFlo™ Accurate Pump PCS Tubing

AcuFlo™ accurate pump PCS Tubing is specially designed for biopharmaceutical applications where accuratedispensing and consistent flow rates are required.

AcuFlo™ tubing has well finished and extremely smooth fluid contact surface that helps in consistent flow rates without any particle generation.

  • Accurate Dispensing: AcuFlo™ accurate pump PCS tubing comply with USP<1151> PHARMACEUTICAL DOSAGE FORMS
  • No Drip Formation: AcuFlo™ accurate pump PCS Tubing do not show any drip formation at different fill volumes with a pause time of 30 seconds with water.
  • No Spallation : No Particle Generation: AcuFlo™ accurate pump PCS Tubing comply with USP <788> for Particulate Matter in Injections and USP <790> for Visible Particulate in Injections over 4 hours of pumping.
  • Consistent Flow Rates: AcuFlo™ accurate pump PCS Tubing offers consistent flow rates.
Single use disposable filling lines for pumping and accurate fluid transfer from surge bags to filling needles
Operational
Sterilization By Irradiation up to 50 kGy
Shelf Life MDI PCS tubing has a shelf life of 5 years.
Assurance
Bioburden Bioburden level is < 1000 cfu/device as per ISO 1737-1:2018.
Fiber Release Passes test as per USP and comply with USFDA Title 21 CFR Part 210.3(b)(6) for fiber release.
Particle Release Complies with USP <788> test for particulate matter in injections.
Indirect Food Additive These PCS tubing meet the FDA Indirect Food Additive requirements cited in 21 CFR 177.1520.
Endotoxin Testing Aqueous extracts exhibit <0.25 EU/ml as established by Limulus Amebocyte Lysate (LAL) Test as per USP <85>.
Non Animal Origin The materials used for manufacture of mdi tubing do not contain substance of animal origin as per EMEA/410/01 Rev. 03.
Extractables Extractable profiles for PCS tubing are available as per BioPhorum Best Practices Guide for Extractable Testing of Polyemric Single Use Components used in Biopharmaceutical Manufacturing.
Sterility The gamma sterilization process for gamma sterilized tubing has been validated as per ISO 11137 to ensure a sterility assurance level (SAL) of 10-6.
Package Integrity MDI tubing with 15 M roll length are packed in double polyethylene bags to ensure package integrity during transit as well as to prevent particulate contamination.
Biosafety Passes Biological Reactivity test, In-Vivo, as per USP<88> for Class VI plastics. Passes the Biological Reactivity Tests, In Vitro for Cytotoxicity as described in USP<87>.

To order, please specify the catalog number and quantity.

Example: AFSA016048AXX01, 2 rolls