FilterCheck® 06 : Integrity Test Equipment
MDI FilterCheck® 06 is a 21 CFR compliant, automatic, compact microprocessor controlled integrity test equipment designed for integrity testing of membrane, cartridge and capsule filters.
This state of the art equipment combines the latest mass flow measurement technology with embedded intelligence systems to provide very high precision and accuracy.
It also includes additional intelligence in the form of an on board computer with capacitive touch screen and specially developed, user-friendly software with simple menu driven data input.
Latest Technology for High Precision and Accuracy
- Direct Mass Flow Measurement
- Independent of Temperature and Pressure Variation
Compatibility with Sterile Processing Areas
- Electropolished SS316L Surface
- No Instrument Cooling Fan
- Wireless and Paperless to minimize particle generation in sterile areas
Easy Portability
- Smallest Integrity Tester on the market
- Weight: 9kg
All types of Integrity tests
- For Hydrophilic as well as Hydrophobic Filters
- Option for Quick Test for Higher Productivity
21 CFR part 11 Compliance for Data Reliability and Security
- Three Access levels with Password Security
- System Audit Trail
- Online repository
- FilterCheck™ 06 web view
Self Test
- Internal communication check
- Inlet pressure check
- Pressure verification check
- Internal leakage check
Self Diagnosis
- Internal Leakages
- External Leakages to ensure Test Filter Connectivity
Shows Real Time Test Status
- Graphical representation of test results
- PDF report generation
Large Storage Space >10GB for
- Test Templates
- Test Reports
Inbuilt Equipment Safety with separateHydrophobic Filters for
- Entrained moisture in Inlet Air
- Inadvertent back flow of test fluids
Features | |
Stainless steel body | SS316L |
Dimension (Main test Unit) | 35 x 32 x 12 cm |
Weight | 9Kg |
Voltage | 100-240V AC |
Input Frequency | 50Hz |
Power | 75W |
Battery Pack | 14.4 V, 4.5 AH |
Capacitive Touchscreen | 7″ |
Fuse | 4A |
Battery Charger (Inbuilt) | Yes |
Power Backup | 3 hours |
External Wireless Printer | Yes |
Charging time | 2 hours |
Operating System | |
Operating Range | |
Maximum Inlet Pressure | 100 psi |
Relative Humidity | 10 – 80% |
Storage Temperature | 5° – 35°C |
Operating Temperature | 8° – 30°C |
Measuring Range | |
Diffusion Flow | 0.2 – 100 ml/min |
Bubble Point | 15 – 80 psi |
Water Intrusion Rate | 0.03 – 10 ml/min |
Accuracy | |
Diffusion Flow | + 5% or 1ml whichever is greater |
Bubble point | + 0.7 psi |
Water Intrusion Rate | + 5% or 0.02ml whichever is greater |
Support Documents | |
Certificate of Quality | Yes |
Calibration Report | Yes |
Installation Qualification | Yes |
OQ (Test Equipment) | Yes |
OQ (Application Software) | Yes |
Performance Qualification | Yes |
File Type | Download |
---|---|
Brochure | Download |